7.1.1
Emergency lid release
In the event of a power failure, it is possible to manually open the
centrifuge lid.
• Switch off the mains power switch and disconnect the power cord from
the socket.
• Remove the plug (see figure, item 1) from the opening on the left side of
the control panel, e.g. with a screwdriver.
Fig. 31: Position of the opening for the emergency lid release
• Insert the supplied hexagon socket key horizontally into the hole. The
key will be guided through a funnel-shaped tube to the shaft of the lid
lock motor.
Fig. 32: The emergency lid release key must be inserted horizontally (similar to illustration)
• Unlock the motorised lid lock with the supplied hexagon socket key
(size 5) by turning it anti-clockwise.
• Then, reinsert the plug.
WARNING
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.