![Sigma 146774 Operating Manual Download Page 58](http://html1.mh-extra.com/html/sigma/146774/146774_operating-manual_1267883058.webp)
Refrigerated Centrifuge with Heating System Sigma 3-30KHS
7 Malfunctions and error correction
58 / 91
Version 09/2014, Rev. 1.15 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0701402
Pos : 168 /100 Sigma/110 BA Z entrif ugen Sigma (Proj ekt e)/ 3-30KS_3- 30KHS/070 Stör ungen und F ehl ersuche/ 070-0010- 0010 N ot entriegel ung 3-30KS_KH S @ 29\m od_1405321917640_68.docx @ 198051 @ 3 @ 1
7.1.1
Emergency lid release
In the event of a power failure, it is possible to manually open the centrifuge
lid.
• Switch off the mains power switch and disconnect the power cord from
the socket.
• Remove the plug (see figure, item 1) from the opening on the left side of
the control panel, e.g. with a screwdriver.
Fig. 30: Position of the opening for the emergency lid release
• Insert the supplied hexagon socket key horizontally into the hole. The
key will be guided through a funnel-shaped tube to the shaft of the lid
lock motor.
Fig. 31: The emergency lid release key must be inserted horizontally
(similar to illustration)
• Unlock the motorised lid lock with the supplied hexagon socket key by
turning it anti-clockwise.
• Then, reinsert the plug.
WARNING
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.
Pos : 169 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1