Refrigerated Centrifuge with Heating System Sigma 3-30KHS
6 Using the centrifuge
Version 09/2014, Rev. 1.15 of 08/05/2020
• sb
31 / 91
Translation of the original operating manual, part no. 0701402
6.2.2.3
Installing a rotor with the g-lock
®
system
The handling of a rotor with the g-lock
®
system is described in a separate
instruction manual.
NOTE
See the separate instruction manual concerning the g-lock
®
rotor quick-
lock system!
Pos : 114 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 115 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0031 Eins et zen v on Z ubehör 4plätzig 3-30KS_KH S @ 27\mod_1405319206727_68.docx @ 192521 @ 4 @ 1
→
6.2.2.4
Installation of accessories
• Only use vessels that are suitable for the rotor.
• In the case of swing-out rotors, all of the compartments must be
equipped with buckets.
• Always load opposite compartments of the rotors with the same
accessories and fill to avoid imbalance.
Centrifugation with vessels of various sizes
Working with vessel of various sizes is possible. In this case, however, it is
very important that the rotationally symmetrical inserts are identical (see
figure).
correct
incorrect
Fig. 7: Permissible and impermissible loading of a swing-out rotor with vessels of various
sizes (example illustration)