![Sigma 146774 Operating Manual Download Page 25](http://html1.mh-extra.com/html/sigma/146774/146774_operating-manual_1267883025.webp)
Refrigerated Centrifuge with Heating System Sigma 3-30KHS
4 Storage and transport
Version 09/2014, Rev. 1.15 of 08/05/2020
• sb
25 / 91
Translation of the original operating manual, part no. 0701402
Pos : 86 / 100 Sigma/ 110 BA Zentrif ugen Sigma (Pr ojekte) /3- 30KS_3-30KH S/ 040 Lagerung und Tr ans port/ 040-0040 Verpack ung 3-30KS_KH S @ 29\mod_1405321883281_68. doc x @ 197953 @ 2 @ 1
4.4
Packaging
The centrifuge is packaged in a wooden crate.
• After taking off the lid, remove the side panels.
• Remove the packaging material.
• Lift the centrifuge upwards with a lifting device to lift it safely. When
lifting the centrifuge, always reach under the centrifuge from the side.
CAUTION
The centrifuge weighs approx. 100 kg!
• Retain the packaging for any possible future transport of the centrifuge.
Pos : 87 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 88 / 100 Sigma/ 110 BA Zentrif ugen Sigma (Pr ojekte) /3- 30KS_3-30KH S/ 040 Lagerung und Tr ans port/ 040-0050 Trans portsic herung 3- 30KS_KHS - Übersc hrift @ 29\mod_1405321884012_68. doc x @ 197967 @ 2 @ 1
4.5
Transport safety device
Pos : 89 / 100 Sigma/ 110 BA Zentrif ugen Sigma (Pr ojekte) /3- 30KS_3-30KH S/ 040 Lagerung und Tr ans port/ 040-0051 Trans portsic herung 3- 30KS_KHS - Text @ 36\mod_1434605997246_68. doc x @ 263932 @ @ 1
The transport safety device consists of a screw which is located at the
bottom panel.
CAUTION
The transport safety device must be removed prior to start-up because the
screw locks the motor bearings!
Removal
• Lift the centrifuge upwards at the front side. Always reach under from
the side.
• Put a suitable object, e.g. a wooden block, between the table and centri-
fuge. The locking screw can now be seen at the bottom panel.
1
Locking screw
Fig. 5: Locking screw (transport safety device)
• Remove the transport safety screw by turning it anti-clockwise by way of
the rotor wrench (included in the scope of supply) or by way of an open-
end wrench or ring wrench (AF10).
• Retain the transport safety device for the possibility of the return of the
centrifuge.
Pos : 90 / 010 U nivers almodule/ Absc hnit tsw echsel @ 0\m od_1202124514062_0.docx @ 418 @ @ 1
Pos : 91 / 010 U nivers almodule/ Seitenw ec hsel @ 0\m od_1202116244312_0. doc x @ 105 @ @ 1