User Manual 17
3.1.5
Signal Track
Turn on or off the signal tracking function. It is used to track the signal whose frequency is unstable
and the instantaneous change of the amplitude is less than 3 dB. By marking cursor 1 on the
measured signal, the change of the measured signal can be tracked and measured all the time.
The signal tracking process is shown in the following figure:
Figure 3-1 Signal Tracking Flow
⚫
When Marker1 is on, turn on signal tracking, a point whose amplitude does not change more
than 3 dB near Marker1 will be searched and marked, and the frequency at that point will be
set to center frequency.
⚫
When Maker1 is off, turn on signal tracking, Marker1 will be activated, and a peak search will
be performed, and then the peak frequency will be set to center frequency.
⚫
Signal tracking function is only available in sweep analysis. Signal tracking function is turned
off in following cases:
Zero Span mode
Traces are not updated, including single sweep mode or View mode.
Cont Peak function turned on
Other non-SA measurement modes