User Manual 9
1.9
Firmware Operation
1.9.1
Check System Information
Users can get the system information by press
System
> “
System
” ->“
About
”, including
⚫
Product Model, Serial and Host ID
⚫
Software Version and hardware Version
⚫
Option Information
1.9.2
Load Option
Refer to the procedures below to activate the options you have purchased.
1
Press
System
> “System”-> “Load Option”
2
Enter the license key in the onscreen window. Press
Enter
to confirm your input and
terminate the license key input. Or
3
Load the .lic file provided by pressing
File
> “
Load
” from internal memory or USB stick.
The option will be enabled after rebooting.
1.9.3
Firmware Upgrade
Follow this procedure to update the instrument firmware:
1
Download the firmware package from an official SIGLENT website.
2
Extract and copy the .ADS file into the root directory of an USB stick.
3
Plug the USB stick into the USB Host connector. Press
System
> “
System
”-> “
Update
”; find
the .ADS file in USB stick.
4
Press the “
Load
”, the analyzer will perform the update process automatically.
The upgrade process will take several minutes. When the upgrade is completed, the
machine will reboot.
Any interruption during the update process will result in update failure and system
data loss. This is
not covered under the warranty
and the user will bear repair
costs and shipping.
Do not remove the USB storage device until the update is finished.