4.7.11 Countinuous Peak
Enable or disable continuous peak search. The default is Off. When enabled, the system will
always execute a peak search automatically after each sweep in order to track the signal under
measurement.
Command format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:CPSearch[:STATe]
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:CPSearch[:STATe]?
Instructions
Enable the continuous peak search function
Gets the status of the continuous peak search function switch
Parameter type
Boolean
Parameter Range
OFF|ON|0|1
Return
0|1
Example
:CALCulate:MARKer1:CPSEarch ON
4.7.12 Marker to
4.7.12.1
M->CF
Set the center frequency of the analyzer to the frequency of the current marker.
⚫
If the
Normal
marker is selected, the center frequency will be set to the frequency of the
current marker.
⚫
If the
Delta
or
Delta Pair
marker is selected, the center frequency will be set to the frequency
of the Delta Marker.
Command format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8[:SET]:CENTer
Instructions
Sets/gets the value of the marker X axis to the center frequency .If the marker
is not open, this command automatically turns the marker on the center
frequency.
Example
:CALCulate:MARKer1:CENTer
4.7.12.2
M->Start Freq
Set the start frequency of the analyzer to the frequency of the current marker.
⚫
If the
Normal
marker is selected, the start frequency will be set to the frequency of the current
marker.
⚫
If the
Delta
or
Delta Pair
marker is selected, the start frequency will be set to the frequency of