User Manual 113
Instructions
Sets/gets the attenuation value
Parameter type
Integer
Parameter Range
0 dB ~ 51 dB
Return
Integer
,
unitdB
Example
:POWer:ATTenuation 10
4.4.3
RF Preamp
Control the state of the internal preamplifier (PA) located in the RF input signal path. When the
signal-under-measurement is small, turning on the preamplifier can reduce the displayed average
noise level to aid distinguishing small signals from the noise.
The corresponding icon “
PA
” will appear at the left side of the screen when the preamplifier is
turned on.
Command format
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Instructions
Sets/gets the preset amplifier inside the switch spectrometer
Parameter type
Boolean
Parameter Range
OFF|ON|0|1
Return
0|1
Example
:POWer:GAIN ON
4.4.4
Scale
Set the logarithmic units per vertical grid division on the display. This function is only available
when the scale type is set to “
log
”.
⚫
By changing the scale, the displayed amplitude range is adjusted.
⚫
Current signal amplitude range that can be displayed:
The Minimum range: Reference level –10 × current scale value.
The Maximum range: The reference level.
Command format
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision?
Instructions
Sets/gets the scale on which trace logarithms are displayed
Parameter type
Float
Parameter Range
1 dB ~ 20 dB