Self-Calibration
The self-calibration program can quickly calibrate the oscilloscope to
reach the best working state and the most precise measurement. It is
recommended to perform a self-calibration if the change of ambient
temperature is more than 5
℃
. Make sure the oscilloscope has been
warmed up or operated for more than 30 minutes before the
self-calibration.
Please perform the self-calibration as follows:
1.
Disconnect everything from all inputs.
2.
Operate Utility>Do Self Cal, and it will pop up the following dialog.
Select Continue to start the self-cal program.
Figure 2 CalibrationInterface
3.
The oscilloscope will not respond to any operations until the self-cal
is finished. After the self-cal is completed, touch the screen or press
any button to exit.