Copyright © Sigicom AB 2021
33
5.8.6.
Firmware Upgrade
This section describes how to upgrade to latest FW version when the C22
is in the office. To upgrade a C22 which is monitoring out in the field (so
called
Remote firmware upgrade
), see section 5.10.1.
The latest FW release is available for download from the Sigicom Support
page:
http://support.sigicom.com/infra/index.html
Note! Monitoring must be OFF, and communication mode set to
Flight mode (see section 5.8.2), to perform a firmware upgrade.
This is the procedure to do this:
1.
Set the C22 in USB mode (see section 5.8.5, step 1 to 3).
2.
Copy the new firmware (a .bin-file plus a .meta-file) to the file
system root on the memory card.
3.
Return from USB mode (see section 5.8.5, from step 4 until
reboot).
4.
The copied firmware files will be indicated by “
FW
” in the
upper left corner of the main screen.
5.
Go to the firmware upgrade menu screen:
→ Settings and actions → Firmware upgrade.
6.
Press
and wait for reboot. The C22 instrument will then
restart with the new firmware.