9
2.2.
INFRA D10 Micro Lead Acid battery
•
The short special battery cable included with the INFRA D10
Micro must be used for safety reasons (see chapter 8
Accessories).
•
Remove the battery from the INFRA D10 Micro before
mounting, demounting, carrying or transporting.
•
Use a nominal 12V, max 30Ah sealed Lead-Acid battery with the
INFRA D10 Micro
•
Use battery AGM or Gel type of batteries to minimize risk of
leakage.
•
Only use chargers provided or recommended by Sigicom, see
chapter 8 Accessories.
•
Under no circumstances, a charger which can supply more than
14.7V or 3A may be used
•
Never keep the battery in temperatures above 50 ºC (122 ºF)
•
If battery is shorted, it may burn the operator or generate sparks.
•
When charging the battery, in or out of the INFRA Datalogger,
place all parts in a well-ventilated room, away from open fire or
electric sparks.
•
If misused, the battery may generate inflammable gas, smoke or
leak sulfuric acid.
•
A worn-out battery, a battery with cracks or with signs to be
swollen or leaking should be replaced immediately.
•
The INFRA D10 Micro shall be mounted upright with the battery
standing at the bottom of the enclosure.
•
The battery strap included with the INFRA D10 Micro shall be
used to secure the battery.
•
If an external battery is used, the internal battery must be
disconnected.