PVA-3000 Reference Manual
December 2, 2019
Sifos Technologies
Page
90
4.13. PowerShell Multi-Pair Measurement Utilities
Multi-Pair measurement utilities are provided to automate the sequencing of measurements such as
PSD
,
SNR
,
Wideband Return Loss
(Echo), and
Wideband Crosstalk
(Xtalk) across pairs or pair groups on a given PVA Test
Port. PSD and SNR measurements will be specified for either 100BaseT (2 pairs) or 1000BaseT (4 pairs) while Bulk
Return Loss and Bulk Crosstalk are restricted to just 1000BaseT link-ups.
Command
Slot
Command Parameters
Returned
Parameters
mpair_psd
port
link 100
|
link 1000
<
start
fstart
> <
stop
fstop
> <
avg
nAvgs
>
<
plot
<
color
>>
Slot
,
Port
+
{Pair No +
(
READY
+
Star dB +
En dB +
Max_d
Min_d
Min_dB)
|
UNLINKED
…
Pair No +
(
READY
+
Star dB +
En dB +
Max_d
Min_d
Min_dB)
|
UNLINKED
}
Perform 1000BaseT (4-Pair) or 100BaseT (2-Pair) PSD
measurements and report power levels and frequencies for
Start_Freq, End_Freq, Max_dB, Min_dB. 2-Pair 100BaseT
completed by connecting both MDI and MDI-X. Utility will first
verify presence of all required calibrations.
link 100
Produce 2-pair (MDI & MDI-X) PSD measurements
for 100BaseT link-ups.
link 1000
Produce 4-pair PSD measurements for 1000BaseT
link-ups.
start
fstart
Specifies non-default start frequency for frequency
response analysis. Range is 0.02 MHz to 80 MHz. Default
is 1 MHz.
stop
fstop
Specifies non-default end frequency for frequency
response analysis. Range is 0.2 MHz to 100 MHz. Default
is 100 MHz.
avg
nAvgs
Specifies non-default number of averages to use
in measurement. Range is 1 to 64. Default is 16.
nAvgs
must be between 48 and 64 if
fstart
< 1 MHz.
plot
Produce graphical frequency vs power plots of each
pair’s PSD measurement. Must use PowerShell WISH for
this feature. Will erase any previously displayed PSD plots.
color
assign a non-default graph color scheme of classic, sky,
rain, sand, desert, or mellow. Specify pair to configure
default, pair-specific color scheme.
mpair_snr
port
link 100
|
link 1000
<
avg
nAvgs
>
Slot
,
Port
+
{Pair No +
(
READY
+ SNR)
|
UNLINKED
…
Pair No +
(
READY
+ SNR)
|
UNLINKED
}
+ AVG_SNR
Generate 1000BaseT (4-Pair) or 100BaseT (2-Pair) average
and present individual pair SNR readings. 2-Pair 100BaseT
completed by connecting both MDI and MDI-X.
link 100
Produce 2-pair (MDI & MDI-X) SNR measurements
for 100BaseT link-ups.
link 1000
Produce 4-pair SNR measurements for 1000BaseT
link-ups.
avg
nAvgs
Specifies non-default number of averages to use
in measurement. Range is 1 to 64. Default is 8.
mpair_echo
port
<
avg
nAvgs
>
Slot
,
Port
+
{Pair No +
(
READY
+ Echo)
|
UNLINKED
…
Pair No +
(
READY
+ Echo)
|
UNLINKED
}
+ AVG_Echo
Generate 1000BaseT (4-Pair) average and individual pair
Bulk Return Loss readings. Only available in 1000BaseT link-
ups. Utility will first verify presence of all required
calibrations.
avg
nAvgs
Specifies non-default number of averages to use
in measurement. Range is 1 to 64. Default is 8.
mpair_xtalk
port
<
avg
nAvgs
>
Slot
,
Port
+
{Pair Group +
(
READY
+ Xtalk)
|
UNLINKED
…
Pair Group +
(
READY
+ Xtalk)
|
UNLINKED
}
+ AVG_Xtalk
Generate 1000BaseT (6-PairGroup) average and individual
pair Bulk Crosstalk readings. Only available in 1000BaseT
link-ups. Utility will first verify presence of all required
calibrations.
avg
nAvgs
Specifies non-default number of averages to use
in measurement. Range is 1 to 64. Default is 8.
Summary of Contents for PhyView PVA-3000
Page 22: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 22 ...
Page 70: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 70 ...
Page 108: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 108 ...
Page 124: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 124 ...