51
Type B, C
•
Unscrew the whole flow splitter with a hex key
Type D (G 1/2“)
•
First unscrew the locking pin (underside of the body) with a hex key
•
Unscrew the flow straightener with a suitable tool
•
Pull the flow splitter out of the body
Flow Splitter Assembly
•
Carry out the steps described above in reverse order
•
After correct assembly flush
red-y
with dry inert gas.
•
Check that the cleaned measuring device is functioning correctly by checking the zero
point and some defined measurement values.
Recalibration
It is mandatory to recalibrate the flow measurement of the instrument after replacing the flow
splitter for optimal accuracy.
5.11.
Returns
When returning a
RedyCompact
please insure suitable packaging. Please contact us before you
return the unit to us. To repair the unit we require a detailed and accurate description of the
problem, the required service and the possible causes of the faults.
Instructions for Returning Your Instrument for Service
The following information will help you return your instrument to Sierra Instruments' Factory
Service Center and will ensure that your order is processed promptly. Prices may vary depending
on the flow range, type of gas and operating pressure of your unit. To request detailed pricing,
contact your local Sierra Instruments distributor or
contact one of our offices directly
Please follow these easy steps to return your instrument for factory
service:
1. To obtain a Return Materials Authorization (RMA) number from the Sierra Instruments to
http://www.sierrainstruments.com/rma/new.php
2. Once you have created an account, click on the Submit New RMA tab and fill in the RMA form
and follow the instructions. You will receive an email confirmation once you have submitted
your RMA.
3. Print a copy of the RMA (that now includes RMA #) and send a copy of the RMA form along
with your meter back to the factory.
If you require service beyond calibration, but do not know which service(s) will be required,
describe the symptoms as accurately as possible on the RMA form.
Warning!
If the device has come into contact with aggressive or toxic gases, please ensure that it is
properly cleaned/flushed before returning the device to us. Please always complete the
contamination declaration form when submitting your RMA