Connecting
5.3 Electrical connection
ULTRAMAT 23
76
Manual, 3/2016, A5E37100388-003
Ex analyzers
The following applies to all devices in the hazardous area:
Analyzers envisaged for use in hazardous areas in accordance with CSA Class I Div. 2 must
be provided with a safety bracket which protects the power connector from being
unintentionally disconnected (see arrow in above picture). This bracket is enclosed loose
with the analyzer and must be attached before switching on.