Functions
8.4 Configuration
ULTRAMAT 23
136
Manual, 3/2016, A5E37100388-003
Table 8- 3
Full-scale value of the analog output current for upward limiting of measured value
Selectable
Analog current
(neg. MV on)
Analog current
(neg. MV off)
0 - 20 mA
21.0 mA
21.0 mA
2 - 20 mA
21.0 mA
1)
21.0 mA
1)
4 - 20 mA
21.0 mA
1)
21.0 mA
1)
NAMUR - 20 mA
20.5 mA
20.5 mA
1)
If measuring range 2 is set to the maximum possible value, the limiting value of range 2 ... 20 mA
is at 20.9 mA and of range 4 ... 20 mA at 20.8 mA.
Analog output during FCTRL (analog output during function control)
Note
If a fault is present on the analyzer, only the values of the setting "Output during faults"
apply. Simultaneous occurrence of values of the setting "Output during FCTRL" are ignored
in this case.
The status "FCTRL" (function control) is set
•
During an AUTOCAL procedure
•
During the warm-up phase
•
During a calibration procedure
•
During remote control via the communication interface
•
In the uncoded state:
Output of the following measured values is possible here:
●
Hold: The value measured directly prior to commencement of a function control is output
unchanged. This also applies to the limits which are output (see section Parameters:
●
Actual: The measured value is continuously updated.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...