Functions
8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
135
Analog output 0/2/4/NAM mA (start-of-scale value of analog output)
You can set the following start-of-scale values for the analog
current range in the 3rd line:
•
0 mA
•
2 mA
•
4 mA
•
NAMUR
You can switch suppression of negative measured values on
or off in the 4th line. The "ON" option is preset, and means that
negative measured values can also be output. With a setting of
2 or 4 mA as the lower limit, values below this down to 0 mA
can therefore be output, i.e. negative measured values can be
displayed (live zero).
If the output of negative measured values is switched off
("OFF" position), the current output is limited to the start-of-
scale value.
If 2 or 4 mA is set as the lower limit, the output is now indeed
limited to 2 or 4 mA. In the adjacent display, the start-of-scale
value of the analog current range is set to 4 mA.
The following tables represent the correlation between analog current outputs and
measuring range limits.
Table 8- 1
Start-of-scale value of the analog current output
Selectable
Analog current
(neg. MV on)
Analog current
(neg. MV off)
0 - 20 mA
0 mA
0 mA
2 - 20 mA
2 mA
2 mA
4 - 20 mA
4 mA
4 mA
NAMUR - 20 mA
4 mA
4 mA
Table 8- 2
Start-of-scale value of the analog output current for downward limiting of measured value
Selectable
Analog current
(neg. MV on)
Analog current
(neg. MV off)
0 - 20 mA
0.0 mA
0.0 mA
2 - 20 mA
0.0 mA
2.0 mA
4 - 20 mA
0.0 mA
4.0 mA
NAMUR - 20 mA
3.8 mA
4.0 mA
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...