Installing the Version 1.2 Software Update
38.
Reload the image file from the CD to the system.
1.3 Installing the Version 1.2 System Software Update
Use the following steps to install the version 1.2 system software update.
1.
Insert the MO diskette containing the version 1.2 software in the MO drive.
2.
Press the
F4
key to display the Preset screen.
3.
Use the trackball to scroll down the menu to the
System/Language
menu
item and select that menu item.
4.
Press
SET
.
5.
Select the Retain Current Preset radio button: the system automatically
stores the settings for all existing system presets, QuickSets, OB tables
and formulas, and the current language selection.
6.
Select the Upgrade System button.
7.
The system instructs the user to restart the system: Restart the system by
pressing the green On/Off switch to de-energize the system.
8.
Wait 15 seconds and then press the green On/Off switch to energize the
system.
9.
The system will display a progress bar at the bottom of the screen, as the
system software installation proceeds; the loading of the software requires
approximately 10 minutes.
10.
When the software installation process is completed, the system displays
the active image screen.
11.
Press the MO eject button to remove the diskette.
1.4 Restoring the Service Data
Use the following instructions to restore the service data.
1.
Press the
F4
key and select the
Service
menu item.
2.
Enter the service password and select the Ok button to display the Main
Menu screen.
3.
Select the Backup/Restore button to display the Backup / Restore screen.
4.
Select the Restore radio button located at the top of the screen.
5.
Check the All Packages checkbox to restore all service data.
!
Caution:
Do not attempt to reload the DIMAQ-IP version 1.023 software to a
system that is operating with version 1.2 system software installed. This opera-
tion will fail and display the following message: “Please check the DIMAQ-IP
software version; DIMAQ- IP will shut down soon.”
Note:
If additional disks are required to complete the installation, follow the
instructions displayed on the screen.
Summary of Contents for SONOLINE G50
Page 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Page 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Page 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Page 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Page 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Page 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Page 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Page 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Page 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Page 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...