W
EIGHING
F
UNCTIONS AND
P
ARAMETERS
Name
Type
Address
Default
Range of Values / Meaning
Reference
Operating mode
BYTE
DBB111
0
Operating mode:
0
NAWI filling procedure
1 NAWI
emptying
procedure
2 Force
measurement
3 Belt
scale
4 Loss-in-weight
scale
5
Bulk flow measurement device
Reserve_3_110 BYTE
DBB112
0 Reserved
Load cell type
BYTE
DBB109
0
0: analog load cell
1: METTLER TOLEDO type WM or WMH
2...255 analog load cell
Timeout digital LC
WORD
DBW110
240
Timeout in ms for receiving the cyclic weight
value of the digital load cell
Regulations
STRING
[4]
DBB116
“----“
"OIML" : = Regulation code OIML
"----": = No regulation code
Weight unit
STRING
[4]
DBB122
kg
Weight unit
Weight unit large
STRING
[4]
DBB128
t
Large weight unit
Length unit
STRING
[2]
DBD134
m
Length units (belt scale only)
U Factor
REAL
DBD138
1000
Conversion factor weight unit > Large weight
unit
Determination time 3
TIME
DBD140
10000
Adjustment or zero setting time for dynamic
zero setting, taring or adjusting 0... max. 900
000 ms (belt scale only)
144
Table 5-2
DR3 Allocation
5.2.1
DR3 - Adjustment Digits 0, 1, 2, 3, 4, for the Zero Point and Adjustment Weights 1,
2, 3, 4
The analog measurement value coming from the load cells is converted into a
digital value with an analog-digital converter. A weight value is then calculated from
this digital value. All functions of the SIWAREX FTC then uses this weight value for
executing their task.
To calculate the weight value from the digital value, the characteristic curve of the
measurement system must be determined. In the simplest case, the characteristic
curve is defined by points 0 and 1. The first operating point (point 0) is always
determined by the empty scale alone with its own construction weight. The weight
of its own construction, the scale load cells deliver a measurement voltage to the
SIWAREX FTC. After the analog-digital conversion of the measurement voltage,
the digital value (adjustment digit 0) is assigned the zero point (zero weight).
For determining the average value of the adjustment point for the belt scale (zero
point, adjustment weight) over a predefined period of time, the determination time
must be defined. This time entry is not applicable for other operating modes.
If the scale is loaded with a known calibration weight (e.g. with 50 % of the
measurement range), then the second operation point can be determined. The new
digital value from the analog-digital converter is now assigned the calibration weight.
The characteristic curve can be further adjusted using a maximum of three other
points that must lie above point 1.
SIWAREX FTC
5-35
Summary of Contents for SIWAREX FTC
Page 1: ...SIWAREX FTC Device Manual Status 10 2010 Module for Belt scale ...
Page 12: ......
Page 134: ...11 Empty 11 122 SIWAREX FTC ...