Troubleshooting
7
7.1
Troubleshooting
Operator display screen after power failure or power cycle
Following the commissioning of the SITRANS WM300 MFA, the final display is the overview
of the relay status and process variables. The F4 key or equivalent is pressed, the device will
reset back to the initial start-up display. If the F4 key is pressed, the operator can
sequentially scroll though the setup to return to the run display.
In the event of a power loss or power cycle, the remote display will repower to the last valid
display screen. If the on-board display does not reload to the setup screen, press the "ESC"
key on the base module to return to the last valid display screen.
Notes:
●
If diagnosis does not solve the malfunction, the probe, pre-amp or SITRANS WM300
MFA may be defective.
–
To find out if the probe is defective (non-IMA type only; i.e. MSP-3 and MSP-9):
i. Disconnect probe from pre-amp.
ii. Connect an ohm meter across the black and white leads.
iii. Nominal probe impedances are as follows:
MSP-3 and MSP-9
750 ohms
If impedance deviates substantially from these values, an open or short circuit
condition is indicated.
Summary of Contents for SITRANS WM300 MFA
Page 6: ...Table of contents SITRANS WM300 MFA 6 Operating Instructions 08 2018 A5E45090709 AA ...
Page 10: ...SITRANS WM300 MFA Overview SITRANS WM300 MFA 10 Operating Instructions 08 2018 A5E45090709 AA ...
Page 58: ......
Page 66: ......
Page 80: ...Support C 2 Certificates SITRANS WM300 MFA 80 Operating Instructions 08 2018 A5E45090709 AA ...