Technical specifications
SITRANS TR200/TR300
Operating Instructions, 04/2007, A5E01071992-01
61
Calculation of differential value
2 identical thermocouples (TC)
(TC1 - TC2 or TC2 - TC1)
Response time
≤ 250 ms for 1 sensor with broken wire monitoring
Broken wire monitoring
Can be switched off
Cold junction compensation
Internal: with integrated resistance thermometer Pt100
external: with external Pt100 IEC 60751 (two-wire or
three-wire input)
external fixed: Cold junction temperature can be set as
fixed value
Measuring range
Programmable
Min. measuring span
Min. 50 to 100 °C (90 to 180 °F)
Characteristic curve
Linear to temperature or special characteristic curve
Millivolt transmitter
Measured variable
DC voltage
Sensor type
DC voltage source (DC voltage source is possible via a
resistor that is connected externally)
Units of measurement
mV
Response time
≤ 250 ms for 1 sensor with broken wire monitoring
Broken wire monitoring
Can be switched off
Measuring range
Parameterizable max. -100 to 1100 mV
Min. measuring span
2 mV or 20 mV
Overload capacity of the input
-1.5 VDC to 3.5 VDC
Input resistance
≥ 1 MΩ
Characteristic curve
Linear to voltage or special characteristic curve
Output
Output signal
4 to 20 mA, two-wire line
SITRANS TR300 supports communication additionally
via HART Rev. 5.9
Auxiliary power supply
DC 11 to 35 V (to 30 V for Ex)
Max. load
(U
aux
-11 V)/0.023 A
Overshoot range
3.6 mA to 23 mA , infinitely adjustable
(default range: 3.84 mA to 20.50 mA)
Error signal (e.g. in case of sensor
failure)
3.6 mA to 23 mA , infinitely adjustable
(default value: 22.8 mA)
Sampling cycle
0.25 s
Damping
Software filter 1st order 0 to 30 s (can be set in
parameters)
Security
Against reverse polarity
Summary of Contents for sitrans tr200
Page 2: ......
Page 20: ...Installation SITRANS TR200 TR300 20 Operating Instructions 04 2007 A5E01071992 01 ...
Page 32: ...Commissioning SITRANS TR200 TR300 32 Operating Instructions 04 2007 A5E01071992 01 ...
Page 66: ...Dimension drawings SITRANS TR200 TR300 66 Operating Instructions 04 2007 A5E01071992 01 ...
Page 74: ...Glossary SITRANS TR200 TR300 74 Operating Instructions 04 2007 A5E01071992 01 ...
Page 77: ......