Functions
8.18 Operating hours counter in temperature classes
SITRANS TR200/TR300
Operating Instructions, 04/2007, A5E01071992-01
53
8.18
Operating hours counter in temperature classes
The SITRANS TR200/TR300 has different operating hours counters. The operating hours
counters are for monitoring the connected process flows.
1. Operating hours counter for transmitter electronics
●
Monitors the number of operating hours that the transmitter has been in continuous
operation subject to the ambient temperature.
●
Records the process of operating hours of the transmitter in 9 ranges of ambient
temperature.
●
Begins with the first commissioning at the plant.
●
The user cannot reset or adjust the operating hours counter or temperature ranges.
●
The operating hours counter is only updated when it is in measuring mode. The operating
hours counter is not updated in simulation mode.
2. Operating hours counter for process variable
●
Monitors the process of the sensor connected to the transmitter in different process
ranges.
●
Records the process of operating hours of the process variable in 9 ranges. Subdivision
takes place subject to the connected sensor and its sensor limits. The user cannot adjust
the ranges.
●
The operating hours counter is reset automatically if one of its following parameters is
changed:
–
Sensor class
–
Sensor type
–
Interface connection
–
Sensor connection
–
Sensor factor
The parameter assignment software SIPROM T for SITRANS TR200 or SIMATIC PDM or
HART communicator on SITRANS TR300 can read out the operating hours counters. The
operating hours counters are automatically stored to the non-volatile memory once an hour.
If the power supply to the device is interrupted, the operating hours counters are available
upon restarting the device.
Summary of Contents for sitrans tr200
Page 2: ......
Page 20: ...Installation SITRANS TR200 TR300 20 Operating Instructions 04 2007 A5E01071992 01 ...
Page 32: ...Commissioning SITRANS TR200 TR300 32 Operating Instructions 04 2007 A5E01071992 01 ...
Page 66: ...Dimension drawings SITRANS TR200 TR300 66 Operating Instructions 04 2007 A5E01071992 01 ...
Page 74: ...Glossary SITRANS TR200 TR300 74 Operating Instructions 04 2007 A5E01071992 01 ...
Page 77: ......