Functions
8.16 Factory parameters
SITRANS TR200/TR300
50
Operating Instructions, 04/2007, A5E01071992-01
8.16
Factory parameters
The menu item Device ➜ Factory parameters ➜ Reload factory settings resets the
configuration of the transmitter to its default values.
Note
Factory Reset
•
The menu item Device ➜ Factory parameters ➜ Reload factory settings resets the
configuration of the transmitter to the default values in the table below.
•
This Factory Reset has the same effect on "ex stock devices" as it has on "customer-
specific configured devices".
•
Furthermore, resetting the transmitter to its default values also resets any customer-
specific digital-to-analog trim and sensor calibration (one point trim or two point trim).
Once a factory reset has been performed, the transmitter has the following configuration:
Parameters
Reset to value
TAG
Are not reset.
Description
Are not reset.
Message
Are not reset.
Serial number
Are not reset.
Installation date (electronics)
Are not reset.
Sensor class
Resistance thermometer
Sensor type
Pt100 DIN IEC 751
Interface
Normal connection
Sensor connection
Three-wire input
Sensor factor
1.00
Sensor offset 1
0.00 °C
Zero point
0 °C
End of measurement
100 °C
Unit
°C
Broken wire monitoring
ON
Short-circuit monitoring
OFF
Short circuit limit
1.00 Ω
Lower end point analog output
Are not reset.
Upper end point analog output
Are not reset.
Alarm value
Are not reset.
Linearization type
Linear to temperature
Damping
0.00 s
Operating hours counters PV
Are all reset to 0 h.
Operating hours counters, field device
Are not reset.
Slave pointers PV
Are all reset to 0.
Slave pointers electronics temperature
Are not reset.
Manufacturer data sensor
Are not reset.
Summary of Contents for sitrans tr200
Page 2: ......
Page 20: ...Installation SITRANS TR200 TR300 20 Operating Instructions 04 2007 A5E01071992 01 ...
Page 32: ...Commissioning SITRANS TR200 TR300 32 Operating Instructions 04 2007 A5E01071992 01 ...
Page 66: ...Dimension drawings SITRANS TR200 TR300 66 Operating Instructions 04 2007 A5E01071992 01 ...
Page 74: ...Glossary SITRANS TR200 TR300 74 Operating Instructions 04 2007 A5E01071992 01 ...
Page 77: ......