Operate
6.2 Parameter description
SITRANS P500 with HART
142
Operating Instructions, 12/2014, A5E02344528-09
Purpose:
Changing the PIN
Description:
This parameter allows you to change your numeric PIN. For this you need
your PIN or SUPER USER PIN (4711).
"5.3.5 Disable Protection"
Purpose:
Switch off access protection
Description:
The entire access protection is deactivated. For this you require the SUPER
USER PIN (4711), unless you have already logged on as a SUPER USER.
After the device is switched off, any PIN or SUPER USER PIN (4711) that
was input ceases to be valid.
Every parameter can be read and changed without previous input of a PIN.
6.2.6.4
"5.4 Auto Log Out"
Factory setting:
On
Adjustment range:
•
Off
•
On
Purpose:
Reset time
Description:
This parameter defines whether the device returns to write-protected status
on its own, following the defined time period of 5 minutes. Any user pass-
word that may be entered will cease to be valid and will have to be reen-
tered, if necessary.
6.2.7
"6 Language"
Factory setting:
English
Setting range:
•
English
•
German
•
French
•
Italian
•
Spanish
Purpose:
Selection of the language
Description:
This parameter allows you to select the language for the menu.
The following language can be ordered as an optional package:
•
Chinese
•
Japanese
•
Russian