Operate
6.2 Parameter description
SITRANS P500 with HART
118
Operating Instructions, 12/2014, A5E02344528-09
"2.5.5 Measvaluedisplay"
Factory setting:
PV unit
Setting range:
•
PV unit
•
%
•
mA
Purpose:
Selection of the PV display
Description:
This parameter allows you to select whether you wish to display the main
indication value as a digital measured variable in the respective unit as a
percentage output value or as a current value in mA.
"2.5.6 Backlight"
Factory setting:
Off
Setting range
•
Off
•
On
Purpose:
Switching background lighting on / off
Description:
This parameter allows you to switch the background lighting of the display
on and off. The intensity of lighting depends on the voltage and output cur-
rent.
"2.5.7 Contrast"
Purpose:
Setting the display contrast
Description:
This parameter allows you to set the contrast of the display.
"2.5.8 Bargraph"
Factory setting:
Off
Adjustment range:
•
Off
•
On
Purpose:
Activate / deactivate the horizontal bar chart
Description:
This parameter allows you to activate a horizontal bar chart which shows the
measured value of the PV.
Note:
You cannot simultaneously display the second, third and fourth display val-
ues.