Technical data
13
13.1
Input
13.1.1
Gauge pressure
Gauge pressure input
Measured variable
Gauge pressure
Measuring span (continuous‐
ly adjustable) or measuring
range, max. operating pres‐
sure (in accordance with
2014/68/EU Pressure Equip‐
ment Directive) and max. test
pressure (in accordance with
DIN 16086) (for oxygen
measurement, max. 100 bar
and 60 °C ambient tempera‐
ture/process temperature)
Measuring span
1)
Maximum permissible
operating pressure MAWP
(PS)
Maximum test
pressure
8.3 … 250 mbar
0.83 ... 25 kPa
0.12 … 3.6 psi
4 bar
0.4 MPa
58 psi
6 bar
0.6 MPa
87 psi
0.01 … 1 bar
1 ... 100 kPa
0.15 … 14.5 psi
6 bar
0.6 MPa
87 psi
9 bar
0.9 MPa
130 psi
0.04 … 4 bar
4 ... 400 kPa
0.58 … 58 psi
20 bar
2 MPa
290 psi
30 bar
3 MPa
435 psi
0.16 … 16 bar
0.016 ... 1.6 MPa
2.3 … 232 psi
45 bar
4.5 MPa
652 psi
70 bar
7 MPa
1015 psi
0.63 … 63 bar
0.063 ... 6.3 MPa
9.1 … 914 psi
80 bar
8 MPa
1160 psi
120 bar
12 MPa
1740 psi
1.6 … 160 bar
0.16 ... 16 MPa
23 … 2321 psi
240 bar
24 MPa
3480 psi
360 bar
36 MPa
5221 psi
4 … 400 bar
0.4 ... 40 MPa
58 … 5802 psi
400 bar
40 MPa
5802 psi
600 bar
60 MPa
8702 psi
7 … 700 bar
0.7 ... 70 MPa
102 ... 10153 psi
800 bar
80 MPa
11603 psi
800 bar
80 MPa
11603 psi
Gauge pressure measuring limits
Low measuring limit
2)
● Measuring cell with silicone oil filling
30 mbar a/3 kPa a/0.44 psi a
SITRANS P320/P420 with 4 to 20 mA/HART
Operating Instructions, 11/2019, A5E44852162-AC
199