Diagnosing and troubleshooting
10.3 General fault codes
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 08/2014, A5E32221386-AC
175
Code/
Icon
Meaning
Corrective Action
S: 31
* Flash error.
Repair required: contact your local Siemens
representative.
S: 32
IDENT number conflict.
Ensure value of the Ident number selector is correct for
the network configuration. If it is correct, the device needs
to be re parameterized by the PLC.
S: 33
* Factory calibration for the internal temperature
sensor has been lost.
Repair required: contact your local Siemens
representative.
S: 34
* Factory calibration for the device has been lost. Repair required: contact your local Siemens
representative.
S: 35
* Factory calibration for the device has been lost. Repair required: contact your local Siemens
representative.
S: 36
* Unable to start microwave module.
Repair required: contact your local Siemens
representative.
S: 37
* Measurement hardware problem.
Repair required: contact your local Siemens
representative.
S: 38
* Microwave module hardware failure: unable to
calculate distance measurement.
Repair required: contact your local Siemens
representative.
S: 43
* Factory calibration for the radar receiver has
been lost.
Repair required: contact your local Siemens
representative.