7.8
Analog Input Setup
Note
Some versions of the FUP1010 portable flowmeter have the ability to read analog input
signals. The [Analog Inp Setup] menu cell is not displayed if the operating system of your
model does not include this facility.
The optional Analog Input Setup function assigns an active analog input to a measurement
channel/path. The flowmeter provides two DC current (4-20mA) and two voltage (0-10V) input
ports for single channel, Dual Channel and Dual Path units. The DC current input ranges from
a zero level of 4 mA to a full scale of 20 mA and allows you to enable this port and then span
it to any desired scaling.
For example, when using the analog input Aux function, assigned numeric variables for
temperature monitoring might be spanned as follows: 4mA=1 (0°C / 32°F) and 20mA=100
(100°C / 212°F). Temperature data sent to the Aux 1 input port will feed through the flowmeter
and report the scaling data to the Datalogger.
Although the flowmeter receives these signals, the standard operating system will not use
them as a vaiable in any of its operations. The analog input only acts as a convenient "feed-
through" to add data to Datalogger reports.
Note
There will be no numeric units shown next to the output data. Be aware of the appropriate
units when entering scaling data and when viewing the resultant output on the Datalogger
report.
Note
Refer to the Installation Drawings or I/O Module markings for the locations of these inputs
and wiring procedures.
I/O Data Control
Analog Inp Setup
Iin1
Input
Off
Aux
4 ma
Numeric entry
20 ma
Numeric entry
Iin2
Same as Iin1
Vin1
Input
Off
Aux
0 volts
Numeric entry
10 volts
Numeric entry
Vin2
Same as Vin1
The flowmeter recognizes the first analog input variable that is assigned to any given parameter
and ignores any subsequent input with the same assignment.
Functions
7.8 Analog Input Setup
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
89
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......