Functions
7.13 Resistive Temperature Device (RTD) Calibration
FUH1010 IP65 NEMA 4X Precision Volume
124
Operating Instructions, 01/2013, A5E32122003-AA
Ice Bath RTD Calibration
Use deionized water and ice mixture at 0°C (32°F) equilibrium for an ice bath. Ensure
temperature with a reference thermometer. Siemens can not assume responsibility for the
incorrect design, construction or operation of an Ice Bath.
NOTICE
Sensor Damage
If RTD sensor makes direct contact with ice during an ice bath calibration procedure the
sensor may be damaged and the calibration results will be incorrect.
Do not allow an RTD sensor to make direct contact with ice during an ice bath calibration
procedure.
To perform a 0°C (32°F) calibration:
1.
Immerse RTD sensor in deionized water and ice mixture. Stir the mixture constantly.
2.
In the [RTD Calibrate] menu move the highlight by pressing the <Up/Down Arrow> to the
RTD you want to calibrate (RTD 1 or RTD 2).
3.
To access the RTD option list press <the Right Arrow>. Move the highlight to [User Cal]
then press <ENTER>. This triggers the pop-up window.
4.
After the RTD sensor reaches equilibrium at 0°C (32°F), press <ENTER> to recalibrate
the RTD sensor.
5.
To verify the calibrated reading, go to Dual Path Flow [Diagnostic Data/Liquid Data] menu
to check the current RTD output. Make sure that it coincides with the reading of the
reference thermometer. Repeat for the other RTD, if necessary.
Summary of Contents for SITRANS FUH1010
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ...21614 C 1011NFPS 7 ...
Page 232: ...1011NFPS 7 C 21614 ...
Page 233: ...1011NFPS 7 C 21614 ...
Page 234: ...21614 C 1011NFPS 7 ...
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...21614 C ...
Page 241: ......
Page 242: ......
Page 243: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 244: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 245: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 246: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 263: ...Index FUH1010 IP65 NEMA 4X Precision Volume 202 Operating Instructions 01 2013 A5E32122003 AA ...