
Functions
7.12 Analog Output Trim
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA
121
Voltage Output Trim (Vo1 & Vo2)
Note
Can be trimmed to within .0025 V of nominal.
To calculate a voltage output:
1.
Set up a multimeter to read volts, then connect it to the supply and return terminals of the
voltage output under test.
2.
Move the highlight to the port to be tested by pressing the <Up/Down Arrow>. Then press
the<Right Arrow> and then press <Down Arrow> to move the cursor to [Trim @ 2V].
$QDORJ2XW7ULP
6LHPHQV
'XDO3DWK>@3DWK
7ULP9ROWRXWSXW
7ULP,R
2SHUDWH
7ULP,R
2SHUDWH
7ULP9R
!2SHUDWH
7ULP9R
7ULP#9
7ULP3JHQ
2SHUDWH
7ULP3JHQ
2SHUDWH
3.
Press <ENTER>. This triggers a 2.00 Volts pop-up window. The multimeter should now
be reading 2.00 Volts.
4.
If the multimeter reading does not match, use the numeric keys to type in the voltage
reading.
5.
Press <ENTER> to register setting. This adjusts the flow meter’s DAC (digital-to-analog
converter) so that a 2.00 Volts output corresponds with 2.00 Volts on the multimeter.
6.
Re-check the multimeter to make sure that it is now reading 2.00 Volts.
Summary of Contents for SITRANS FUH1010
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ...21614 C 1011NFPS 7 ...
Page 232: ...1011NFPS 7 C 21614 ...
Page 233: ...1011NFPS 7 C 21614 ...
Page 234: ...21614 C 1011NFPS 7 ...
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...21614 C ...
Page 241: ......
Page 242: ......
Page 243: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 244: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 245: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 246: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 263: ...Index FUH1010 IP65 NEMA 4X Precision Volume 202 Operating Instructions 01 2013 A5E32122003 AA ...