Modbus holding registers
A.3 Modbus holding registers tables
FC410 with Modbus
Operating Instructions, 02/2016, A5E33120874-AC
133
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value range Access
level
530
Unsigned /
2
Modbus Parity
Framing
RS 485 parity and framing
8 data bits are always used
0 = even parity, 1 stop bit
1 = odd parity, 1 stop bit
2 = no parity, 2 stop bits
0
0 to 2
Read /
Write
600
Unsigned /
2
Restart
communication
Restart Modbus
communication
Write:
•
0 = No effect
•
1 = Restart
Read:
•
Always 0
-
0 to 1
Write
A.3.7
Characteristics
Table A- 14 Sensor
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value range Access
level
2113
Float / 4
Minimum Frame
Temperature
Lower limit of the frame
temperature
-50 [°C]
Read only
2115
Float / 4
Maximum Frame
Temperature
Lower limit of the frame
temperature
200 [°C]
Read only
4043
String / 16 Sensor size
Nominal sensor diameter
(DN)
-
-
Read only
4053
String / 16 Hazardous area
approval
Hazardous area approval of
the sensor
-
-
Read only
4078
String / 16 Wetted materials
Sensor enclosure material
-
-
Read Only
Table A- 15 Volumeflow calibration
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value range Access
level
2103
Float / 4
Maximum Volume-
flow Capacity
Maximum volumeflow
measurement capacity of
the sensor
Sensor size
specific [m
3
/s]
1)
0 to 0.177
Read only
1)
: See Sensor dimension dependent default settings (Page 139).