10.2
Application Problems
Table 10-1
Troubleshooting list
Error
Diagnosis
Remedy
Sensor does not work or no
display or no output signal
Power supply missing
Connect or switch on power supply
Device fuse defective
Check device fuse and replace if necessary
Transmitter does not react to
controls
Reflections in the glass
Clean glass
Glass is not tight to the operating unit
Screw in the electronic cover tightly
Transmitter cannot be para‐
meterized by the operating
module
Personal code is activated
Enter personal code (menu 6.1 or deactivate by
entering 0 in menu 6.2)
Local operation lock activated by HART or
PROFIBUS
Cancel operation lock
Output signal OK but no dis‐
play visible
Ambient temperature outside permissible
range
Remote design: bring transmitter into a tempera‐
ture range of 0 to 50 °C
Display defective
Change operating module
Flow display = 0 at available
flow
Low flow cut-off set too high
Reduce low flow cut (menu 3.1.6)
Flow opposite to preferable direction at set‐
ting of measuring direction to "forwards
only"
Switch over preferred direction menu (3.1.5.1) or
set measuring direction to "forw.+rev." (menu
3.1.5.2)
Measured value set to 0 via digital input
Switch over signal at digital input (menu 4.3.2) or
switch off function (menu 3.3.1)
Cable connection to sensor defective (only
for remote design)
Check cable connection and exchange as required
SmartPLUG defective
Have SmartPLUG exchanged (only by service per‐
sonnel)
Transmitter indicates flow at
zero flow
Hydraulic zero point drift, application-de‐
pendent
Perform zero correction (menu 6.6.2) or enter
correction value directly (menu 6.6.1)
Grounding problems
Provide potential equalization (see Installation
with grounding rings (Page 28))
Fluctuating measuring re‐
sults at constant flow
Air, gas occlusions
Eliminate gas occlusions or use gas trap
Grounding problems
Provide potential equalization (see Installation
with grounding rings (Page 28))
Vibrations in the sensor and/or in the cable
run
Fix the sensor better, strap cables
Medium conductivity too low
Check conductivity; measuring method may be
unsuitable
Proportion of solids too high
Reduce proportion of solids, measuring method
may be unsuitable
Very unstable measurement, strongly fluc‐
tuating values, output going towards zero
Check for moisture in SmartPlug
If fitted with round SmartPlug call factory product
support
Troubleshooting
10.2 Application Problems
TRANSMAG 2 with MAG 911/E
80
Operating Instructions, 03/2021, A5E00102775-AC
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...