4.4
Turning the local display
The local display can be turned in 90° steps to enable better reading in case of vertical installation
or overhead assembly.
1. Switch off the power supply.
2. Release the catch on the lid of the electronics compartment with a 3 mm Allen key.
3. Unscrew the cover.
4. Carefully release the fastening hooks of the local display using a screwdriver or similar tool
5. Pull out the unit, turn it to the desired position and push it back in.
6. Screw the lid back on and mount the lid catch.
①
Fastening hooks
Figure 4-19
Unlocking the fastening hooks on the local display
Installing/mounting
4.4 Turning the local display
TRANSMAG 2 with MAG 911/E
32
Operating Instructions, 03/2021, A5E00102775-AC
Summary of Contents for SITRANS F TRANSMAG 2