Modbus communication
A.9 Modbus holding registers tables
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
257
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
9440
Float / 4
Fail-safe value Output value in case of
a fault and when Fail-
safe behavior (9439) is
configured to 4.
0.0
[Hz]
0.0 - 12500.0
Read /
write
7163
Float / 4
Forced value
Forced value when
forcing is active.
1.0
[Hz]
0.0 - 12500.0
Read /
write
Channel 4 pulse output settings
Table A- 63 Channel 4 pulse output settings
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
9460
Unsigned / 2 Process value
Select process value
for the pulse output.
0
•
0: Mass flow
•
1: Volume flow
•
27: Standard volume
flow rate (only hydro-
carbon and gas ap-
plications)
Read /
write
9462
Unsigned / 2 Flow direction
Flow direction filter.
0
•
0: Positive direction
•
1: negative direction
•
2: Positive and nega-
tive directions
Read /
write
9465
Float / 4
Amount of vol-
ume
Amount value to gen-
erate an output pulse.
Applicable if process
value (9460) is set to 1.
0.001
[m
3
]
(8993)
-
Read /
write
9463
Float / 4
Amount of
mass
Amount value to gen-
erate an output pulse.
Applicable if process
value (9460) is set to 0.
1.0
[kg]
(8992)
-
Read /
write
9467
Float / 4
Amount of
standard vol-
ume
Amount value to gen-
erate an output pulse.
Applicable if process
value (9460) is set to
27.
1.0
[Sl]
(8994)
-
Read /
write
10255
Unsigned / 2 Pulses per
amount
Number of pulses
which will be output
when configured
amount is reached.
1
-
Read /
write
9474
Float / 4
Pulse width
Pulse duration.
0.1
[s]
0.00004 - 4.0
Read /
write