Technical data
11.3 Accuracy
MAG 6000 I / MAG 6000 I Ex de transmitter
64
Operating Instructions, 04/2016, A5E02083319-AG
11.3
Accuracy
For accuracy reference conditions, see below.
Image 11-2 MAG 6000 I / MAG 6000 I Ex de with MAG 1100 (not PFA), MAG 1100 HT, MAG 1100 F
(not PFA), MAG 5100 W, MAG 3100 P, MAG 3100 and MAG 3100 HT
Reference conditions
(ISO 9104 and DIN/EN 29104)
A calibration certificate is shipped with every sensor and calibration data are stored in
SENSORPROM
®
memory unit.
Medium temperature
20 °C ± 10 K (68°F ± 18 °F)
Ambient temperature
25 °C ± 10 K (68°F ± 18 °F)
Supply voltage
U
n
± 1%
Warming-up time
30 minutes
Incorporation in conductive pipe section
Inlet section
Outlet section
10 x DN (DN ≤ 1200/48")
5 x DN (DN > 1200/48")
5 x DN (DN ≤ 1200/48")
3 x DN (DN > 1200/48")
Flow conditions
Developed flow profile
Reference conditions for sensor calibration