C.2.3
200-299
Param‐
eter ID
Meter
version
Display
view
Parameter/data type
Factory settings
Data range
Fixed parameter or meter data that are not changeable
200
All
O4
Fault status
No faults
0 to 8191, binary presented with
information 1 for bit 0
1: Insulation error
2: Coil current error
3: Amplifier overload
4: Database checksum error
5: Low power warning
6: Flow overflow warning
7: Pulse A overload warning
8: Pulse B overload warning
9: Consumption interval warning
10/L: Leakage warning
11/E: Empty pipe warning
12/C: Low impedance (high con‐
ductivity) warning
13/d: High flow limit warning
14/A: Reverse flow warning
If no error/warning is pending,
the message is "No faults"
201
All
-
Alarm configuration list
254 = Alarm 2 to 8 enabled 0 to 8191, See 200
If no alarm is enabled, the mes‐
sage is "None enabled"
202
All
-
Date of fault log reset
PS3 production date and
time
year-month-day T hours:mi‐
nutes:seconds
203
All
O4
Non optimal measure time, in‐
cluding the time of device run‐
ning with insulation test fault,
coil current fault, amplifier
fault, database checksum
fault, flow overload fault, or
empty pipe fault.
0
204
All
-
Reset the fault log and faults
No
Yes / No
Value is reset to "No" when up‐
loading device parameters to pc
after command execution
205
All
-
Call up acknowledge
No
Yes / No
Value is reset to "No" when up‐
loading device parameters to pc
after command execution
206
All
-
Battery alarm level
10%
0 to 100%
Appendix
C.2 Parameter lists
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE
139