Parameter lists
A
The following tables show the various parameters available via PDM.
The parameters are accessible at three different levels:
● Read: Maintenance level
● Read/Write (R/W): Specialist level (write access is only permitted after entering the user
password; default password is 1000)
● Hardware locked (HW key): The parameter is only accessible when a HW key is mounted
Note
For type-approved and verified FUE380 flowmeter the settings are HW key protected and
therefore read only. This HW key is protected via the verification sealing. The verification
sealing can only be opened by the user with the acceptance of the local authorities.
A.1
Identification
Table A-1
Identification parameters (FW 2.03 and EDD 1.02.08-01)
Modbus
Address
Parameter
Default val‐
ue
Value range
Access
level
Description
40049
1
Application identifier
Identity
R/W
Customer application identification in‐
formation (max 16 characters)
40057
2
Application location
Location
R/W
Customer application location informa‐
tion (max 16 characters)
Device
40032
5
Sensor size
Product de‐
pendent
R/W
Pipe diameter of the sensor. It is only
a text field for measurement system in‐
formation. The value has no influence
on the measurement. Please see inner
diameter at pipe date in parameter 306
40141
8
Totalizer unit
m
3
R/W
Totalizer unit as text for volume.
Change the unit in parameter table if
the device menu "Unit guide (offline)"
does not list the wanted unit.
Note: The device display can show
"m
3
" unit only - other units cannot be
shown on the display, but will be used
and shown online via PDM
FUS380/FUE380
Operating Instructions, 09/2016, A5E00730100-AB
85