Every 5 minutes the advanced power management system of the transmitter calculates the
battery consumption and the remaining battery operation time.
For battery versions, only the internal battery pack has a nominal capacity of 34 Ah giving a
typical operation time up to 6 years in a revenue application.
The ambient temperature of the transmitter also influences the battery capacity.
The typical operation time up to 6 years is based on only 80% battery capacity, low pulse
output frequency, seldom use of communication and an operation time/temperature profile of
5% at 0 °C (32 °F), 80% at 15 °C (59 °F) and 15% at 50 °C (122 °F).
The effect of other temperatures is shown in the figure below.
A variation in temperature from 15 °C to 55 °C (59 °F to 131 °F) reduces the capacity by 17%
(in the figure from 15 Ah to 12½ Ah).
$YDLODEOHFDSDFLW\
7HPSHUDWXUH
&
DS
DFL
W\
WR
9
$K
Note
The installation orientation of the battery pack may influence the battery capacity. Optimal
battery capacity is achieved with the battery pack in an upright position.
10.3
Sensor for FUS380 and FUE380
Table 10-3
Technical data
Description
Specification
Pipe design
2-track sensor with flanges and integrated transducers wet-calibrated from
factory
Nominal size
DN 50, 65, 80, 100, 125, 150, 200, 250, 300, 350, 400, 500, 600, 700, 800,
900, 1000, 1200
Pressure rate
PN 16, PN 25, PN 40
EN 1092-1 flanges:
● type 01: DN 100 to DN 125
● type 11: DN 150 to DN 1200
● type 11 'design': DN 50 to DN 80
Technical data
10.3 Sensor for FUS380 and FUE380
FUS380/FUE380
Operating Instructions, 09/2016, A5E00730100-AB