Technical data
10
10.1
FUS380 and FUE380 systems
Table 10-1
Technical data SITRANS FUS380 and FUE380
Description
Specification
Enclosure design/material
Fiber glass reinforced polyamide in light-gray color
Enclosure
IP67 according to EN 60529 and DIN 40050 (NEMA 4X/6)
Wall mounting kit
Terminal box made of fiber glass reinforced polyamide in light-gray
color with 2 + 2 cable glands (output, supply and transducer cables)
and a stainless steel bracket for wall or pipe mounting
Cable glands
In wall mounting kit:
● 4 x M20 PA plastic cable glands (2 for mains and output cable,
2 for the transducer coaxial cables)
Ambient temperature
-10 °C to +55 °C (14 °F to 131 °F)
1)
Storage temperature
−40 °C to + 85 °C (−40 °F to +185 °F)
Transducer/sensor cables
● Coaxial cable, impedance 75 Ohm, ∅ app. 6 mm;
● Remote version cable length: Max. 30 meter between transmitter
and transducer (5 m (16.4 ft), 10 m (32.8 ft), 20 m (65.6 ft), or 30
m (98.4 ft) available).
Mechanical vibration
2 g, 1 to 800 Hz sinusoidal in all directions according to IEC 68-2-6
Power supply
Battery
2)
:
● Replacable 3.6 V LiSOCI (Lithium Thionyl Chloride)
● Double battery pack 34 Ah or single backup battery 17 Ah;
● Battery exchange interval: 4.2 (6) years at 60°C (140°F)
operation
Mains:
● 87 to 265 V AC (50 to 60 Hz)
Display
● LCD, 8 digits, additional 2 digits and symbols for status
information
● Units: Volume unit: m
3
(default display unit), Flow unit: m
3
/h
(default display unit)
● Alarm codes (F 1 to F 9) for: path 1 or 2 not measuring, internal
failure, power supply failure, flow overload, pulse output
frequency overload, data logger warning
Push button
One push button for toggling between display information
Measuring function
● 0.5 Hz (battery mode)
● 15 Hz (mains-powered)
FUS380/FUE380
Operating Instructions, 09/2016, A5E00730100-AB
79