View process values
The system is now ready for normal operation.
1. Select [View → Process Variables] to see all measured process values.
2. Verify that the fields show the expected values.
Note
Measured Track Values
The Measured Track Values should show stable values within the normal range, i.e. the gains
should be stable values between 3 and 12 (smaller sizes low values and larger sizes high
values) and the actual velocities for the paths should be stable, constant and smoothly
changing values between 0 and 10 m/s.
Store settings on the PC
Store the device settings after verifying the values.
● Choose: [Device → Upload to PG/PC] to get all the settings.
● Store the complete settings on your PC via [File → Export].