6.4.4
Configuring the device
In the following it is described how to configure the device by defining all sensor specific
parameters.
Note
Measurement accuracy
A correct flowmeter installation is required to ensure optimum measurement accuracy.
Read all parameters
Before any parameterization is done it is necessary to read all parameters from the device into
the offline table of SIMATIC PDM. The offline table merely contains default data.
1. Open PDM device driver.
2. Select Upload to PG/PC .. and click [Start] to read all parameters to the offline table.
After closing the dialog box, all loaded parameters should show
(Loaded) in the status
of the PDM table; except date and time formats, they show
(Changed).
3. Store factory set default values on your local PC [File → Export] in order to be able to retrieve
default settings.
Read, write device data
Only parameters (data) shown with white background can be changed.
Note
Status field
●
indicates changed off-line data not yet stored in the device.
●
indicates actual data downloaded to the device.
Commissioning
6.4 Commissioning via PDM
FUS380/FUE380
Operating Instructions, 09/2016, A5E00730100-AB
53