
Operating
7.2 Operating the FST030
SITRANS F Ultrasonic flowmeters FST030 (HART)
Operating Instructions, 08/2017, A5E35971532-AC
75
7.2.4
Handling alarms
When the alarm list is displayed in the main view, press to get more detailed information
about the active alarms.
Table 7- 6
Key functions - alarms list view
Key
Function
Exit alarm list view
Select the item above in the list; keep pressing the key to accelerate scrolling up the
selection list
Select the item below in the list; keep pressing the key to accelerate scrolling down the
selection list
View more information on the selected alarm
Press to exit the detailed alarm information.
Alarm acknowledgement
There are two ways to have the alarms removed from the alarm list.
●
Manual: The alarm remains in the alarm list until the alarm is manually acknowledged
(ack.). Before the alarm can be acknowledged, the cause must be eliminated. Press