4.10.5.2
Density
Density settings
Table 4-13
Density settings
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting options
Access
level
2127
Float / 4
Empty Tube
Limit
Define threshold value
of
empty tube
500 [kg/m
3
]
14 000 to
+14 000
Read /
write
2129
Unsigned /
2
Empty Tube De‐
tection
Set automatic detec‐
tion of
Empty Tube On/Off
0 = off (Empty tube is
off).
1 = on (a density value
below
Empty Tube Limit trig‐
gers an
alarm. All flow rate val‐
ues
are forced to zero %).
0
0 to1
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
59