Channel 3 output current
Table 4-98
Channel 3 output current
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
9125
Unsigned / 2
Simulation
Enables or disables
simulation of channel 3
output current.
0
● 0: Disabled
● 1: Enabled
Read /
write
9126
Float / 4
Simulated value Current value that is
output on the channel
when simulation is ena‐
bled.
0
[mA]
0 - 25
Read /
write
Channel 3 output status
Table 4-99
Channel 3 output status
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
9164
Unsigned / 2
Simulation
Enables or disables
simulation of channel 3
output status.
0
● 0: Disabled
● 1: Enabled
Read /
write
9165
Unsigned / 2
Simulated value Signal level that is out‐
put on the channel
when simulation is ena‐
bled.
0
0, 1
Read /
write
Channel 3 output frequency
Table 4-100 Channel 3 output frequency
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
9205
Unsigned / 2
Simulation
Enables or disables
simulation of channel 3
output frequency.
0
● 0: Disabled
● 1: Enabled
Read /
write
9206
Float / 4
Simulated value Frequency value that is
output on the channel
when simulation is ena‐
bled.
1.0
[Hz]
0.0 - 12500.0
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
179