
Process value 4
Table 4-89
Process value 4
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
11203
Unsigned / 2
Process value
Process value config‐
ured to be monitored
Disabled
● 3000: Mass flow
● 3002: Volume flow
● 3004: Density
● 3010: Medium
temperature
● 3023: Frame
temperature
● 7950: Corrected
volume flow
● 7871: Fraction A
● 7881: Fraction B
● 7800: Fraction A in %
● 7810: Fraction B in %
● 65535: Disabled
Read /
write
11334
Float / 4
Maximum value Drag indicator of maxi‐
mum value specified by
Process value (11203)
Min. floating
point value
[units depend‐
ing on proc‐
ess value]
-
Read
only
11336
String / 32
Maximum value
timestamp
Shows timestamp of re‐
corded maximum value
specified by Process
value (11203)
-
Read
only
11316
Float / 4
Minimum value Drag indicator of mini‐
mum value specified by
Process value (11203)
Max. floating
point value
[units depend‐
ing on proc‐
ess value]
-
Read
only
11318
String / 32
Minimum value
timestamp
Shows timestamp of re‐
corded minimum value
specified by Process
value (11203)
-
Read
only
11207
Unsigned / 2
Reset logging
Command to reset the
drag pointers to the
process value's default
value.
-
-
Write on‐
ly
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
173