4.10.6.1
Identification
System with FCT030 transmitter
Table 4-60
System
Modbus
register
Data type /
Size in
bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting op‐
tions
Access
level
6592
String / 12
Manufacturer
Device manufacturer.
SIEMENS
-
Read only
6104
String / 32
Product name
Product name.
Also shown on the device
nameplate.
SITRANS F
-
Read only
6020
String / 16
Version
Product version according to
order.
-
-
Read only
6120
String / 32
System order num‐
ber
System order number part 1
(MLFB).
Also shown on the device
nameplate.
-
-
Read only
7236
String / 32
System order num‐
ber
System order number part 2
(MLFB).
Also shown on the device
nameplate.
-
-
Read only
7268
String / 32
System order num‐
ber
System order number part 3
(MLFB).
Also shown on the device
nameplate.
-
-
Read only
6088
Unsigned /
4
Final assembly
number
Final assembly number.
Can be used to identify de‐
vice upgrades.
0
-
Read /
write
6576
String / 32
Product serial
number
Unique alpha numerical seri‐
al number. Also shown on
the device name plate.
-
-
Read only
6334
String / 16
Hardware version
System hardware version.
Also shown on the device
name plate.
-
-
Read /
write
(Expert)
6030
String / 16
Firmware version
Product firmware version. Al‐
so shown on the device
name plate.
-
-
Read /
write
(Expert)
8120
String / 32
Long TAG
Unique TAG name.
-
-
Read /
write
8176
String / 16
Descriptor
Description of the measuring
point.
-
-
Read /
write
8136
String / 32
Message
Additional information.
-
-
Read /
write
8152
String / 32
Location
Device location in the plant. -
-
Read /
write
8168
String / 16
Installation date
Installation date of the de‐
vice.
-
-
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
136