Functions
7.10 Analog Output Trim
FUE1010 IP65 NEMA 4X
Operating Instructions, 12/2014, A5E03086491-AC
125
To calculate a voltage output:
1.
Set up the multi-meter to read volts, then connect it to the supply and return terminals of
the voltage output under test.
2.
Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
Arrow>, then press <Down Arrow> to move the cursor to [Trim @ 2V].
3.
Press <ENTER>. This triggers a 2.00 Volts pop-up window. The multi-meter will now be
reading 2.00 Volts.
4.
If the multi-meter reading does not match, use the numeric keys to type in the multi-meter
reading.
5.
Press <ENTER> to register setting. This adjusts the flow meter’s DAC (digital-to-analog
converter) so that a 2.00 Volts output corresponds with 2.00 Volts on the multi-meter.
6.
Re-check the multi-meter to make sure that it is now reading 2.00 Volts.
Pgen Output Trim
Note
Can be trimmed to within 1.25 Hz of nominal.
To calibrate a pulse rate output
1.
Connect a frequency counter to the signal and reference terminals of the pulse rate
output under test.
2.
Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
Arrow>, then press the <Down Arrow> to move the cursor to Trim @ 1kHz.
3.
Press <ENTER>. This triggers a 1 kHz pop-up window. The frequency counter will now
be reading 1 kHz.
4.
If the frequency counter reading does not match, use the numeric keys to type in the
frequency counter reading.
Summary of Contents for SITRANS 7ME3500
Page 2: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ...21614 C 1011NFPS 7 ...
Page 255: ...1011NFPS 7 C 21614 ...
Page 256: ...1011NFPS 7 C 21614 ...
Page 257: ...21614 C 1011NFPS 7 ...
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ...21614 C ...
Page 264: ......
Page 265: ......
Page 266: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 267: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 268: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 269: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...21614 C STEP 4 BAND LENGTH TABLE PERMANENT INSTALLATION SERVICE NOTE STEP 2 STEP 3 STEP 1 ...
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 282: ......
Page 290: ...Glossary FUE1010 IP65 NEMA 4X 216 Operating Instructions 12 2014 A5E03086491 AC ...
Page 294: ...Index FUE1010 IP65 NEMA 4X 220 Operating Instructions 12 2014 A5E03086491 AC ...
Page 295: ......