
Adjustments / Programs
5 - 25
Siemens AG
SPR2-230.061.01
Page 25 of 30
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Evaluation
5
Calculate the deviation between the measured area dose product and the displayed area-
dose product:
The deviation must be < 0.3 .
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg) ]
/ (measured area dose product FDPg)
For deviations > 0.3, the area dose product measurement device must be replaced.
Fig. 8
Fig. 9
Fig. 10
Meas. chamber
Sirephos
I.I.
Center cross
Sirephos
I.I.