Medical Solutions
5 - 18
Adjustments / Programs
Save values and end adjustment
5
•
In the "Available Functions:" window select "INIT OFF & STORE VALUES" and click on
"Execute". Close the status window by clicking on the "OK" button.
For the query "Do you want to reset the Unit", click on "Yes".
Concluding tasks
5
•
Before exiting the service program, save the new parameter values to disk
(Service PC, Data menu, Backup).
Correction of the 0° position of the slot diaphragm display
5
•
Select the "Adjustments", "Parameter" menu at the service PC.
•
Select "Screen display Slot diaphragm" in the Combo box.
•
Click "Get from Unit".
•
Select "Rotation 0 degree position".
•
The parameter in the "Value Actual" field can be changed according to the minimum and
maximum values.
•
Then click "Set Value".
•
Click "Put to Unit".
•
After the SIREMOBIL Iso-C has booted, compare the 0° position of the slot diaphragm
blades to the display on the monitor. If necessary, repeat the programming.
•
Click "Save to file".
•
Leave the service program with "Logoff" and "Quit" to save the parameters as a new file
to disk.
Correct only if the slot diaphragms are shown at an angle of rota-
tion different from that displayed on the monitor. A correction of
+ 10° is possible. Camera rotation and slot diaphragm have to be
set correctly.
The service PC is connected to the serial interface of the
SIREMOBIL Iso-C.
The service program is started.
NOTE