
Adjustments / Programs
5 - 17
Siemens AG
SPR2-230.061.01
Page 17 of 30
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
•
In the window "Available Functions:" select "Slot Diaphragm Open Pos - I.I. Full Format"
and click "Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No" .
The slot diaphragm is closed automatically.
Full format, slot diaphragm closed
5
•
Apply fluoro briefly.
•
Open and close the slot diaphragm display via the keys.
•
Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
•
In the window "Available Functions:" select "Slot Diaphragm Close Pos - I.I. Full Format"
and click "Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No".
The slot diaphragm remains closed.
Zoom format, slot diaphragm closed
5
•
The zoom format is selected automatically.
•
Apply fluoro briefly.
•
Open and close the slot diaphragm display via the keys.
•
Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
•
In the window "Available Functions:" select "Slot Diaphragm Close Pos - I.I. Zoom For-
mat" and click "Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No". The slot dia-
phragm opens automatically (zoom format).
Zoom format, slot diaphragm opened
5
•
Apply fluoro briefly.
•
Open and close the slot diaphragm display via the keys.
•
Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
•
I
n the window "Available Functions:" select "Slot Diaphragm Open Pos - I.I. Zoom For-
mat" and click "Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No".
Enable circle mask
5
•
In the "Available Functions:" window select "Enable Circle Mask" and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click on "No".