2 - 1
Siemens AG
SPR2-230.061.01
Page 1 of 2
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Line voltage adjustment
2
•
Open the back cover of the monitor trolley.
•
Measure the on-site line voltage and line frequency.
•
Adjust transformer T1 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
•
Adjust transformer T2 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
•
Check fuses F1 and F2 to see if they correspond with the values on the label of the
switch-on assembly and replace them, if necessary.
•
Mark the correct line voltage and line frequency on the line voltage label at the back of
the monitor trolley.
Checking the line voltage on the monitor trolley
2
•
Jumper X10 for the basic system must be disconnected.
•
The monitor, memory device and other devices, if present, such as multiformat camera,
video printer or video recorder must be disconnected from the line voltage supply.
Check the secondary voltages for line voltage transformer T1 according to table 1.
Electrical voltage!
See Chapter 1, Safety Information.
Disconnect the SIREMOBIL Iso-C line voltage plug.
from test point
to test point
Voltage
Comment
Transformer T1.31
Transformer T1.32
190 V ~ to
205 V ~
cannot be adjusted
Transformer T1.31
Transformer T1.33
230 V ~ to
246 V ~
cannot be adjusted
Line voltage cable
MEMOSKOP, N
Line voltage cable
MEMOSKOP, L
230 V ~ to
246 V ~
cannot be adjusted
Line voltage cable
Monitor 1, N
Line voltage cable
Monitor 1, L
230 V ~ to
246 V ~
cannot be adjusted
Option:
Line voltage cable
Monitor 2, N
Line voltage cable
Monitor 2, L
230 V ~ to
246 V ~
cannot be adjusted
Option:
Line voltage cable
MULTISPOT, N
Line voltage cable
MULTISPOT, L
230 V ~ to
246 V ~
cannot be adjusted
Option:
Line voltage cable
Video printer, N
Line voltage cable
Video printer, L
230 V ~ to
246 V ~
cannot be adjusted
Option:
Line voltage cable
Video recorder, N
Line voltage cable
Video recorder, L
230 V ~ to
246 V ~
cannot be adjusted
Tab. 1
WARNING